10
Artis zee
Floor-mounted system for surgical angiography
Operating modes
Fluoroscopy
Digital pulsed fluoroscopy, with 10, 15, 30 P/s in 1k/12-bit matrix and digital real-time filtering for advanced noise
reduction with motion detector
Additional fluoroscopy pulse rates from 0.5 to 7.5 p/s* (CAREVISION)
Roadmapping (requires DSA option) with automatic pixel shift
Overlay fade, online superimposing of active fluoro and reference image
Storage of fluoroscopy images, incl. during fluoroscopy
Last Image Hold (LIH)
Fluoro Loop*
Storage and display of dynamic fluoro sequences
The maximum fluoro time that can be saved depends on the pulse frequency selected, e.g., 17 s at 30 p/s, 34 s at 15 p/s
Roadmap Plus*
Simultaneous display of subtracted and native fluoroscopy images
DR – 0.5 - 7.5 f/s*
Individual and serial images in original matrix size, full format and zoom 1 + 2 in 2k*, digital real-time filtering, individual
image and series frame rates from 0.5 f/s to 7.5 f/s native, including time-controlled and manually variable frame rates
Acquisition, display and storage in original matrix size (up to 2k)*
DSA – 0.5 - 7.5 f/s*
Digital subtraction angiography in original matrix size, full format and zoom 1 + 2 in 2k*, digital real-time filtering,
individual image and series frame rates from 0.5 f/s to 7.5 f/s, including time-controlled and manually variable frame rates
Acquisition, display and storage in original matrix size (up to 2k)*
Remask, peak opacification for iodine contrast (MaxOpac) and CO
2
contrast (MinOpac), display of anatomical background
(Landmark) from 0 to 100%
High-speed acquisition for DSA, DR*
Increase of image frequencies for DSA and DR to 10/15/30 f/s
Storage in 1k matrix
Cardiac acquisition*
Acquisition at 7.5, 10, 15 and 30 f/s, acquisition, display and storage in 1k matrix, 12-bit
Cardiac acquisition includes IC Stent* software for enhanced stent visibility: operable at tableside, available in < 30 s
Pediatric option* with 60 f/s (only for 20 x 20 detector)
* Option (2k only with FD 30 x 40)