D-Link DSA-3200 Manuel d'utilisateur Page 28

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 52
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 27
24
VARIABLE XSCANTRIG
VALID VALUES 0 or 1
DEFAULT VALUE 0
DATA TYPE integer
DESCRIPTION Sets the External Trigger as the Frame Trigger.
0 = the internal clock is the frame trigger
1 = the external trigger is the frame trigger
VARIABLE Z C
VALID VALUES 0 or 1
DEFAULT VALUE 1
DATA TYPE integer
DESCRIPTION Sets zero correction on or off.
0 = no correction
1 = zero correction
NOTE 1: ZC must be set to 0 for the following conditions:
1. Model is set to 3207
2. The module has gauge transducers installed.
NOTE 2: ZC must be set to 0 for the following conditions:
1. Model is set to 3207
2. Absolute sensors are installed and the CALB command will
not be used.
NOTE 3: ZC must be set to 1 for the following conditions:
1. Model is set to 3217 or 3218
2. Differential transducers are installed
NOTE 4: ZC must be set to 1 for the following conditions:
1. Model is set to 3207
2. Absolute sensors are installed and the CALB command will be
used.
Vue de la page 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 51 52

Commentaires sur ces manuels

Pas de commentaire